Fringe Correction Tables¶
Here you will find what are the detector-filter combinations that requires a Processed Fringe Frame for the data reduction. Below are one table for GMOS-N and one table for GMOS-S. Each row of these tables corresponds to one of the detectors used in the instrument during its life-time. The five columns in the right contains the broadband filters used in imaging mode. The intersection of the detector rows with the filter columns contains cell with the following information:
- Yes: Requires a Processed Fringe Frame for data reduction;
- No: Does not require a Processed Fringe Frame for data reduction;
- —: Does not have data with these filters.
GMOS-N | i’ | CaT | Z | z’ | Y | |
---|---|---|---|---|---|---|
EEV CCDs | Aug 2001 - Nov 2011 | Yes | Yes | — | Yes | — |
E2V DD CCDs | Nov 2011 - Feb 2017 | Yes | Yes | Yes | Yes | Yes |
Hamamatsu CCDs | February 2017 - Present | No | No | No | Yes | Yes |
More: GMOS-N Fringe Information
GMOS-S | i’ | CaT | Z | z’ | Y | |
---|---|---|---|---|---|---|
EEV CCDs | Commissioning - June 2014 | Yes | Yes | — | Yes | — |
Hamamatsu CCDs | June 2014 - Present | No | No | No | Yes | Yes |